RY-III A Anawsthetic workstation
Inspiratory flow: (6--66)L/min
Adjustable range of tidal volume: (50!1500)mL

Frequency: (6!99) times/min
I: E: 3: 1!1: 9(inspiratory time: 0.3s!6.0s)
Alarm for upper limit pressure: (1.0!6.0)kpa (no les than lowest limit)
Alarm for lowest limit pressure: (0.4!4.0)kpa (lasts 4s---15s)
Maxi safe pressure: Exhaust pressure of pressure-limiting valve: (5.1!6.0)kap

Alarm for lowest ventilation: Alarm will be given within 90s when expiratory flow drops below minimal setting value.
Range of tidal volume of bellows: Adult: (300-1500)mL, child: (50-300)mL.
Alarm for power supply broke off: Warning will be given when power supply is broke off and lasts no less than 120s.

Alarm of gas supply cut off: Alarm for lowest limit pressures will be sounded when gas supply is cut off.
Silence time fo alarm: No more than 120s.